PARAMETERS
|
STANDARDS’ REQUIREMENTS
|
MAHIP SALT SPRAY APPARATUS
|
COMPETITIVE CHAMBERS
|
Temperature Control
|
Of Chamber Atmosphere
|
Chamber Atmosphere
|
Salt Solution
|
Precipitation Rate of Fog
|
1-2 ml/80sq.cm./hr
|
1-2 ml/80sq.cm./hr throughout
|
No measurement/ Control At best, point control
|
pH of Sprayed Sol.
|
Consistently in Range - 6.5-7.2/ 3.1-3.3
|
Consistently in Range
|
At best, erratic control
|
Humidifier
|
Necessary. With Continuous Level Control
|
Present With Continuous Level Control
|
Usually no humidifier present
|
Heater-Electrical
|
Not permitted inside the Chamber
|
Not present inside Chamber
|
No Immersion/ air heaters present inside the chamber
|
Construction
|
No metal permitted inside Chamber
|
No metal present inside Chamber
|
SS Spray nozzles, etc. present inside Chamber.
|
Capacity
|
15c.ft. / 400 Lt minimum
|
431 Lt/ 15.2 c.ft.
|
150-200 Lt.
|
Consistency of Results
|
Consistent throughout the year
|
Consistent throughout the year
|
Inconsistent even during day & night.
|
Recycling of Spray Solution
|
Solution once sprayed should be disposed off
|
No recycling
|
Recycling
|
Fog Venting
|
To be provided
|
Provided
|
Not provided.
|
Digital Temp Display Logging & Control
|
-
|
Provided
|
Usually Not provided
|
Integrated Castor mounted Model
|
-
|
Fully integrated
|
Not integrated
|
Direct Impingement of Spray
|
Not Permitted
|
Continuous Deflected Super Fine Fog/Mist
|
Intermittent Direct spray of Solution droplets.
|
Electrical Safety Features
|
-
|
Provided
|
Not Provided
|
Failure prevention tech.
|
-
|
Provided
|
Not Provided
|
Test validity check (audit) facility
|
-
|
Provided thru proprietary software
|
Not possible
|
Dependence on trained operator
|
-
|
Low
|
High
|